Sindhu.G, Dr. Vamsi Lavu, Dr. Ram Sabarish, Prof. R.Suresh Comparison of the root surface temperature rises following intracanal irradiation with 810nm and 940 nm diode lasers at varied power settings and a standardized environment- An in vitro study. International Journal of Scientific Research, Vol : 2, Issue : 8 August 2013
This study was performed to determine the influence of diode laser wavelength on root surface temperature
rise in a standardized environment.
A total of fifteen freshly extracted teeth were collected and biomechanical preparation was performed. Intracanal irradiation was
performed with diode lasers (810nm & 940 nm) at different power settings (1.5W, 2W, 3W) with an exposure time of 20 seconds in a
wet environment. The temperature rise at the external root surface was assessed using a thermocouple.
A significant rise of temperature was observed for both wavelengths and this temperature rise was above 10 degrees when irradiation was done for 20 seconds
Our study results demonstrate that irrespective of the wavelength of the diode laser (810nm/940 nm) used, the rise in temperature at
the external root surface following intra-canal irradiation was similar.