Original Research Paper Health care

CBCT Motion Tracking Internal Motion Tracking and Validation for Abdominal Radiation Therapy

Gopal Subramanian

Abstract

<p> Purpose To demonstrate that during pretreatment Cone–Beam CT (CBCT) can accurately measure tumor motion and that this method should be</p>
<div>used to validate margins during treatment.</div>
<div>Methods and Materials For 26 patients with abdominal tumors and implanted fiducial markers, tumor motion was measured daily with CBCT and</div>
<div>fluoroscopy for 186 treatment fractions. Fiducial tracking and maximum–likelihood algorithms extracted 3–dimensional fiducial trajectories from</div>
<div>CBCT projections. The daily internal margin (IM) (ie, range of fiducial motion) was calculated for CBCT and fluoroscopy as the 5th–95th</div>
<div>percentiles of displacement in each cardinal direction. We validated CBCT fiducial tracking as an accurate predictive measure of intrafraction</div>
<div>motion by comparing the daily measured IM CBCT with the daily IM.</div>
<div>Results Four–dimensional CT could not accurately predict intrafractional tumor motion for ≥84% of fractions in 96% (IMCBCT ), 94% (IM prefluoro</div>
<div>), and 100% (IM fluoro ) of patients. For patients with median treatment time t < 7.6 minutes, IM CBCT was in agreement with IM fluoro for</div>
<div>94% of fractions (superior–inferior), compared with 67% for the t > 7.8 minutes group, demonstrating the need for patient–specific intratreatment</div>
<div>imaging.</div>
<div>Conclusions Cone–beam CT could replace fluoroscopy for pretreatment verification of simulation IM, reducing patient setup time and imaging</div>
<div>dose. Patients with treatment time t > 7.6 minutes could benefit from the addition of intratreatment imaging.</div>

Keywords: CBCT, Radiotherapy, Marker, Quality Assurance