Dr. Alakesh Choudhury, Dr. Saurabh Borkotoki, Dr. Reetwik kumar Dutta, Study on Blood Glucose levels preanalytically processed on different sample cups over a period of time, INDIAN JOURNAL OF APPLIED RESEARCH : Volume-8 | Issue-2 | February-2018
<p> </p>
<p class="MsoNormal" style="text–align:justify"><span style="font–size:14.0pt;
line–height:115%;font–family:"Times New Roman","serif"">The aim of the study is to compare random blood glucose levels processed in <span style="mso–spacerun:yes"> </span>three different types of collection vials(i.e. Clot,EDTA,NaF) with increase in time.<span style="mso–spacerun:yes"> </span>Blood samples are collected from 30 non–diabetic healthy individuals and are filled in equal amounts separately to clot vials, EDTA vials and NaF vials. After that serum was separated from each vials before estimation by centrifugation over 10 minutes in 3000 rpm and analyzed for blood glucose levels at hourly interval for next 5 hours by a Vitros 250 autoanalyser. The glucose level of the samples tend to fall with increase in time. It falls more rapidly in clot vials than EDTA vials. The fall is minimum in samples collected in NaF vials.<o:p></o:p></span></p>
Keywords: random blood glucose, Clot, EDTA, NaF