Original Research Paper Paediatrics

Comparison of ease of tracheal intubation using Macintosh laryngoscope and AirtraqTM optical laryngoscope in paediatric patients by novice anaesthetists

Dr. Harick Shah Dr.Nandini Dave

Abstract

Method: 64 patients between 2 – 12 years of age requiring general anaesthesia with endotracheal intubation were included. Endotracheal intubation was attempted by a resident anaesthetist under the guidance of a senior anaesthetist. Time taken for intubation, number of intubation attempts, Intubation Difficulty Score, were noted.
Results: Time taken for intubation was 51.594 + 17.544 s in the Airtraq group compared with 64.156 + 20.743 s in the Macintosh group (P =0.011).Using Airtraq 25 subjects were intubated in the first attempt and 7 in the second attempt compared to 17 in the first attempt and 15 in the second attempt for Macintosh (P = 0.035). The intubation difficulty score was 0.469 + 0.803 for Airtraq group versus 1.406 + 1.563 for Macintosh group (P = 0.004).
Conclusion: Airtraq optical laryngoscope is an easy to use device for tracheal intubation, by novice anesthetists, with shorter learning curve. Airtraq reduced the difficulty of intubation and provided better hemodynamic conditions.

Keywords: Airtraq, Macintosh, NoviceAnesthetists