Original Research Paper Science

Development & Validation of Liquid Chromatography Tandem Mass Spectrometry Assay for Quantification of Quetiapine Fumarate in Human Plasma

Dr. Tushar Barot Hardik Brahmbhatt

Abstract

A simple, rapid, sensitive and specific Liquid Chromatography tandem mass spectrometric (LC–MS/MS)
method was developed and validated for the quantification of Quetiapine Fumarate in human plasma.
Analyte was chromatographed on a peerless basic C18 column (50×4.6 mm) 3µm [Chromatopack] with mobile phase
composition of [Acetonitrile: 5mM Ammonium Formate buffer (70:30)] at a flow rate of 0.700 mL/minute and Quetiapine– D4 was used as the internal standard. The assay involves a simple solid–phase extraction procedure of 0.100mL
human plasma and the analysis was performed on a triple–quadrupole tandem mass spectrometer by MRM mode via
electrospray ionization (ESI). The method was linear in the concentration range of 1004.176ng/mL to 5.038ng/mL. The
lower limit of quantification (LLOQ) was 5.038ng/mL.

Keywords: Quetiapine Fumarate, LC–MS/MS, Human Plasma