Eric Rytkin, Computational Fluid Dynamics Analysis of a Novel Stent Geometry for Carotid Artery Stenosis., PARIPEX‾INDIAN JOURNAL OF RESEARCH : Volume-7 | Issue-2 | February-2018
<p> <b style="text–align: justify;">Aim</b></p>
<p class="MsoNormal" style="text–align:justify;line–height:200%">The aim of this research was to perform Computational Fluid Dynamics analysis of a novel auxetic stent.<o:p></o:p></p>
<p class="MsoNormal" style="text–align:justify;line–height:200%"><b>Materials and Methods<o:p></o:p></b></p>
<p class="MsoNormal" style="text–align:justify;line–height:200%">Numerical results of velocity and wall shear stress were retrieved. From that the conclusion of auxetic stent interference on the blood flow was drawn.<o:p></o:p></p>
<p class="MsoNormal" style="text–align:justify;line–height:200%"><b>Results<o:p></o:p></b></p>
<p class="MsoNormal" style="text–align:justify;line–height:200%">Without a stent, minimal wall shear stress was 6.14395e–03 [Pa] and maximal 3.58595 [Pa]. Blood flow velocity was minimum 9.43298e–08 [m/s] and maximum 0.224575 [m/s]. With a stent placed in the internal carotid artery, minimal shear stress was 4.494e–013 [Pa] and maximal 1.695e+001 [Pa]. Blood flow velocity with stent in place was minimum 0 [m/s] and maximum 6.634e–001 [m/s].<o:p></o:p></p>
<p class="MsoNormal" style="text–align:justify;line–height:200%"><b>Conclusion<o:p></o:p></b></p>
<p><span style="font–size:12.0pt;font–family:"Times New Roman","serif";mso–fareast–font–family:
"Times New Roman";color:#222222;mso–ansi–language:EN–US;mso–fareast–language:
EN–US;mso–bidi–language:AR–SA">The auxetic stent has the potential to reduce the rate of restenosis and improve the quality of life among patients with carotid artery stenosis</span></p>
Keywords: Stent, Auxetics, Stenosis, CFD