Volume : IX, Issue : III, March - 2020
Fixation tape: A cause of failure of cuff deflation and pilot assembly malfunction
Dr. Ashish Yadav, Dr. Rupali Arora, Dr. Kirti Kamal, Dr. Monika Bhandari, Dr. Savita Saini, Dr. Geeta Ahlawat
Abstract :
Kinking of the inflation tube of the endotracheal tube can lead to inadequate inflation or failure of deflation of endotracheal tube cuff. We discuss a case of difficult extubation, due to inadequate deflation of the tracheal tube cuff, despite collapse of the pilot balloon, on its aspiration. This was caused by inadvertent kinking of the pilot balloon tubing due to inappropriate tape fixation of the endotracheal tube.
Keywords :
Article:
Download PDF
DOI : 10.36106/ijsr
Cite This Article:
FIXATION TAPE: A CAUSE OF FAILURE OF CUFF DEFLATION AND PILOT ASSEMBLY MALFUNCTION, Dr. Ashish Yadav, Dr. Rupali Arora, Dr. Kirti Kamal, Dr. Monika Bhandari, Dr. Savita Saini, Dr. Geeta Ahlawat INTERNATIONAL JOURNAL OF SCIENTIFIC RESEARCH : Volume-9 | Issue-3 | March-2020
Number of Downloads : 118
FIXATION TAPE: A CAUSE OF FAILURE OF CUFF DEFLATION AND PILOT ASSEMBLY MALFUNCTION, Dr. Ashish Yadav, Dr. Rupali Arora, Dr. Kirti Kamal, Dr. Monika Bhandari, Dr. Savita Saini, Dr. Geeta Ahlawat INTERNATIONAL JOURNAL OF SCIENTIFIC RESEARCH : Volume-9 | Issue-3 | March-2020